-40%
THERMO FISHER UV/VIS VISIBLE SPECTROPHOTOMETER, EVOLUTION EVO 300 PC
$ 1504.8
- Description
- Size Guide
Description
THERMO FISHER UV/VIS SPECTROPHOTOMETER, EVOLUTION EVO 300 PCUntested beyond power up. Assumed to be in excellent, used condition, but sold as is for parts/repair.
The Evolution 300 is ideal for life science, pharmaceutical and QA/QC applications. A wide selection of accessories and a low cost of ownership make it a versatile and affordable addition to the lab. The powerful Xenon lamp-based platform and variable bandwidth make the Evolution 300 ideal for life science, pharmaceutical and QA/QC applications. A selection of VISION software packages expand instrument capability and Smart Accessories make your analyses easier to perform (not included in this auction). The walk-up convenience of the Xenon lamp saves valuable time and increases your lab's productivity.
Product Size
-
Accuracy
±0.004A at 1A, ±0.004A at 2A, ±0.006A at 3AÅ
Accuracy (Photometric)
±0.004A at 1A, ± 0.004A at 2A, ± 0.006A at 3A
Baseline Flatness
±0.0015A (200-800nm), 2.0nm SBW, smoothed
Beam Geometry
Dual-Beam; Quartz Coated
Certifications/Compliance
21 CFR Part 11 and audit-proof IQ/OQ/PQ documentation. ISO 9001:2000
Connections
RS-232
Depth (English)
21 in.
Depth (Metric)
53 cm
Description
Evolution 300 Security UV-Vis Spectrophotometer
Detector Type
Dual Matched Silicon Photodiodes
Drift
<0.0005Abs/hr, 500 nm, 2.0 nm SBW, 2hr warm-up
Electrical Requirements
100/240 V 50/60 Hz
Height (English)
15 in.
Height (Metric)
38 cm
Hertz
50/60 Hz
Includes
Item Description
300 PC
Interface
Computer Control
Lamp
Xenon Flash Lamp
Typical lifetime: >5 years, longer if not using live signal
Warrenty period: 3 year source replacement warrenty
Min. Data Interval
0.5 nm
Monochromator
Modified Ebert
Noise
Photometric:
0A: <0.00018 A
1A: <0.00022 A
2A: <0.00050 A
500 nm, 2.0 nm SBW, RMS
Optical Design
Modified Ebert
Double beam with sample and reference cuvette/accessory positions
Pathlength (Metric)
Up to 100 mm cuvettes
Pharmacopoeia Compliance Testing
Resolution (Toluene in Hexane): Peak/Trough Ratio >2.0
Photometric Accuracy (60mg/L K
2
Cr
2
O
7
): <±0.01 A relative to calibrated sealed standard
Stray Light: <0.13 %T at 198 nm with KCl per EP
220 nm: <0.01%T NaI (behaves identically to KI for USP)
Wavelength Accuracy: ±0.20 nm (546.11 nm Hg emission line), ±0.30 nm 190–900 nm
Wavelength Repeatability: Peak separation of repetitive scanning of Hg line source <0.10 nm
Photometric Accuracy Including Standard Tolerance
1A: ±0.008 A
2A: ±0.010 A
3A: ±0.018 A
Photometric Display
±6 A
Photometric Range
>4 A
Photometric Readout
Absorbance, % Transmittance, % Reflectance, Concentration
Photometric Repeatability
1A: ±0.0025 A
Range (Photometric)
>4A
Resolution (Toluene/Hexane)
Peak/Trough >2.0 at 0.5nm SBW
Scan Ordinate Modes
Absorbance, % Transmittance, % Reflectance, Concentration,
1st-4th Derivative
Scan Speed
1 - 3,800 nm/min
Spectral Bandwidth
Variable 0.5; 1.0; 1.5; 2.0; 4.0 nm
Stray Light
198 nm: 2.9 A KCl
220 nm: 4.2 A NaI
340 nm: 4.3 A NaNO
2
System Requirements
At least Windows XP
Voltage
100/240 V
Wavelength Accuracy
±0.20 nm (546.11 nm Hg emission line); ±0.3 nm for 190 - 900 nm
Wavelength Data Interval
10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05 nm
Wavelength Range
190 to 1100nm
Wavelength Repeatability
Standard deviation of 10 measurements <0.05 nm
Wavelength Scan Speed
3800, 2400, 1200, 600, 240, 120, 60, 30, 10, 5, 1 nm/min Intelliscan
Weight (English)
48.4 lb.
Weight (Metric)
22 kg
Width (English)
24 in.
Width (Metric)
61 cm